Thermocouple Calibration Furnace Parameters

Important technical parameters of the thermocouple calibration furnace:
(1) Detection temperature range 300~1200℃
(2) Furnace length ≥600㎜
(3) The lowest temperature point in the thermocouple calibration furnace does not deviate from the center of the furnace tube by more than 20㎜
(4) Within ±20㎜ of the highest point of the furnace temperature, the temperature gradient is ≤0.4℃/㎝
(5) Power supply voltage: 50HZ 220V±10%
(6) Maximum heating power: 5kVA
(7) Heating time: From room temperature to 800℃, the heating power is not more than 5kVA, which takes about 30 minutes
(8) Shape and size: height × length × width = 1200×700×400
(9) Weight: about 37㎏
Ordering information:
(1) Purchasing a WJL-30 thermocouple calibration furnace and a SKW-1 digital intelligent precision temperature control device can achieve better temperature control results.
(2) This type of thermocouple calibration furnace cannot be used to detect standard platinum-rhodium 30-platinum-rhodium 6 thermocouples and industrial platinum-rhodium 30-platinum-rhodium 6 thermocouples with grade II accuracy. If you want to purchase a calibration furnace for detecting the above-mentioned platinum-rhodium 30-platinum-rhodium 6 thermocouples, our company can arrange production. The detection temperature range is 1000~1550℃, which meets the requirements of the relevant regulations.