Home - Knowledge - Details

How to determine if a hot runner temperature control box needs calibration

Whether a hot runner temperature control box needs calibration should be determined based on a comprehensive assessment of four dimensions: temperature accuracy, process stability, equipment alarm feedback, and service life. The following are systematic judgment criteria, covering technical performance, operational anomalies, and management standards, ensuring scientific and feasible decision-making.

 

I. Excessive deviation between displayed and measured temperature values ​​(most direct criterion)

On-site comparison method: Use a metrology-certified standard temperature measuring instrument (accuracy ±0.1℃) and a thermocouple inserted side-by-side into the same temperature measuring port. Read the value after the system has stabilized for 10 minutes.

Deviation ≤ ±1℃: System normal, no calibration required.

Deviation > ±1℃ and ≤ ±2℃: Calibration recommended to prevent process drift.

Deviation > ±2℃: Calibration required immediately, posing a quality risk.

Example: Set temperature 250℃, temperature control box displays 253℃, standard instrument measures 249℃, deviation +4℃ → Calibration required.

 

II. Abnormal Process Stability (Indirect but critical production signal)

Abnormal Phenomenon

Possible Cause

Does it indicate calibration requirement?

Product weight fluctuation > ±3% under the same mold and parameters

Temperature fluctuation causes melt viscosity change

Strong correlation

Product exhibits short shots, flow marks, silver streaks

Localized insufficient or overheating

Temperature control accuracy needs verification

Uneven color, localized carbonization

Overheating in a certain area, thermocouple inaccuracy

Calibration required

Dimensional variation exceeds tolerance by ±0.1mm per batch

Temperature control lag or drift

High Probability Correlation

Note: If other variables such as raw material batch, injection pressure, and mold wear have been ruled out, temperature runaway is the primary suspect.

 

III. Temperature Control Box Alarms and Abnormal Control Behavior

Alarm Type

Behavioral Characteristics

Calibration Required

Thermocouple Open Circuit

Sensor open circuit or poor contact; calibration is required after replacement

Forced

Frequent Over-Temperature/Under-Temperature Alarms

Actual temperature is not exceeded, but the system reports a false alarm

Temperature Feedback Distortion

Continuous PID Oscillation (above ±5℃)

Unable to stabilize after ruling out parameter setting errors

Abnormal Feedback Signal

Slow Heating Response

Set to 250℃, takes more than 30 minutes to reach the temperature

Sensor Aging or Carbon Deposits

Warning: If alarms are frequent but there is no hardware damage, over 90% of the time it is caused by temperature sensing system drift.

 

IV. Maintenance Cycle and Usage Scenarios Triggering Mechanisms

Usage Scenarios

Recommended Calibration Cycle

Calibration Triggered?

High-Precision Production (Medical, Electronic, Optical Components)

Every 3 Months

Calibrate upon expiration

Regular Injection Molding (Daily Necessities, Packaging)

Every 6–12 Months

Calibrate upon expiration

Multi-Shift Continuous Operation (>18h/day)

Every 3–6 Months

Mandatory Calibration for High-Frequency Use

Equipment Restart After Long-Term Shutdown (>30 Days)

-

Mandatory Calibration

Replacing Thermocouples, Temperature Control Modules, or Molds

-

Recalibration Required

Two Consecutive Calibration Correction Value Changes > ±1.5℃

-

Systematic Drift Exists, Cycle Needs to Be Shortened

 

V. Calibration Record Trend Analysis (Preventative Judgment)

Establish a calibration history archive, recording each correction value (Offset):

If three consecutive calibration correction values ​​show a unidirectional increase (e.g., -0.5℃ → +0.3℃ → +1.2℃), it indicates sensor aging, and the cycle should be shortened to 3 months.

If the correction value remains stable within ±0.2℃, the calibration period can be extended to 18 months, but quarterly spot checks are required.

It is recommended to use an electronic ledger with automatic alerts for approaching calibration dates to avoid human error.

Conclusion: Determining whether a hot runner temperature control chamber needs calibration should not rely on a single indicator, but rather on a four-tiered verification system: measured deviation + process anomaly + alarm feedback + periodic management. Any condition triggering "mandatory calibration" should be implemented immediately to prevent "operation with defects." Regular calibration is not a maintenance cost, but an investment in quality assurance.

info-673-462

Send Inquiry

You Might Also Like